Feature
- Integrated anesthesia workstation with built-in ventilator for smooth and reliable operation
- Advanced safety system with oxygen failure warning, hypoxia protection, and nitrous cut-off
- User-friendly design with clear gas flow indicators, easy controls, and emergency oxygen flush
- Durable, stable construction with stainless steel work surface and ample storage drawers
- Flexible open-to-closed circuit changeover suitable for a wide range of surgical procedures
Technical Specification
Structure: Rigid structure electro-statically powder coated.
Storage Space: Three spacious drawers provided for storage of anesthesia accessories, equipment, etc.
Monitor Shelf: Large monitor shelf at eye level provided.
Cylinder Yokes: Gas-specific pin-indexed yokes for holding 2 Nos. O₂ and 1 No. N₂O cylinders of 5-litre water capacity.
Hypoxia Guard: Ensures that at least 25% oxygen is always maintained in the breathing circuit/gas mixture.
Rotameter: Three-tube rotameter for O₂, N₂O and Air (5-tube cascade type optional).
Non-Return Cum Pressure Relief Valve: Minimizes the risk of backflow and blow-off when pressure exceeds 125 cm H₂O.
Regulators: Separate regulators for each pin-indexed yoke (2 Nos. O₂ and 1 No. N₂O).
Second Stage Regulator: Provided for better safety and stability of the system.
Oxygen Flush: Oxygen flush provided for emergency oxygen (35–40 Ltr/min).
O₂ Failure Warning Device: Self-activated audio warning device to indicate O₂ pressure drop below 2.5 bar.
Nitrous Cut-Off: Ensures N₂O cut-off if oxygen supply falls below 2.5 bar.
Common Gas Outlet (CGO): Conveniently located CGO with ISO standard tapers of 22 mm & 15 mm.
Pipeline Gas Inlets: Non-interchangeable quick-connection pipeline inlets – one each for O₂, N₂O and Air.
Pressure Gauges: Large diameter, color-coded gauges fitted for better visibility and clarity (one each for pin-indexed yokes and corresponding pipeline gas inlets).
Selectatec Manifold: Twin Selectatec type manifold/block for mounting vaporizers.
Table Top: Stainless steel table top for writing and keeping medicines.
Anasthesia Ventilator: Inbuilt / Integrated anesthesia ventilator.
Close Circuit: Closed circuit / Circle absorber single canister (twin canister optional).
Changeover Unit: Easy changeover from open to closed circuit.
Ventilator – Technical Specifications
Tidal Volume: 40 – 1200 mL
Frequency: 8 – 40 bpm
I:E Ratio: 3:1 – 1:3.0
Pressure Limit: 65 cm H₂O
Airway Pressure: 5 – 100 cm H₂O
Trigger: 1 – 10 cm H₂O
Drive Mode: Pneumatically driven and microprocessor controlled
Battery Backup: 90 minutes
Alarms
Low Gas Supply
High Airway Pressure
Disconnection
Low Battery
Anesthesia Vaporizers(ISOFLURANE)
Feature - Calibrated concentration control, high safety
Dimension (L*W*H) - 90 x 185 x 197 mm Millimeter (mm)
Material - Aluminium
Display - Analog concentration scale
Advantage - Precision delivery of gaseous anesthetic agents
Weight - 7.00 Kilogram
Packing
Dimension: 1375 × 600 × 600 mm
Weight: 120 kg (approx.)